Quick Start NIR Workflow: step by step
1.
Check if your NIR-Device Data Format is directly supported (anyway you can convert to JCAMP) :
NIR-Predictor supported Spectral Data File Formats
2.
Download the free NIR-Predictor Software that contains demo data so you can play with it to see if it is the way you want analyse your NIR spectra (no registration needed) :
NIR-Predictor Download
3.
With your "NIR device" measurement software:
- measure samples with NIR, that gets you spectra files,
- label them with a proper sample name, so you know which is which,
- and determine quantitative reference values by Laboratory reference method.
- at least 60 samples with different contents is needed for a minimal calibration.
- NIR-Predictor helps to create a template file (.csv) to enter the Lab values.
4.
Creating your own Calibrations with NIR-Predictor to combine your NIR and Lab data for a calibration request :
watch Video
read Manual
Schnellstart NIR-Workflow: Schritt für Schritt
1.
Überprüfen Sie, ob Ihr NIR-Geräte-Datenformat direkt unterstützt wird (sonst können Sie nach JCAMP konvertieren). :
NIR-Predictor supported Spectral Data File Formats
2.
Laden Sie die kostenlose NIR-Predictor Software herunter, die Demo-Daten enthält, damit Sie mit ihr spielen können, um zu sehen, ob es die Art und Weise ist, wie Sie Ihre NIR-Spektren analysieren wollen (keine Registrierung erforderlich): :
NIR-Predictor Download
3.
Mit Ihrer "NIR Instrument" Messsoftware :
- Proben mit NIR messen, das liefert Ihnen Spektrendateien,
- kennzeichnen Sie sie mit einem richtigen Probennamen, damit Sie wissen, was was was ist.
- Bestimmung der quantitativen Referenzwerte durch die Laborreferenzmethode.
- Für eine Minimalkalibrierung werden mindestens 60 vermessene Proben mit unterschiedlichem Inhalt benötigt.
- Der NIR-Predictor hilft bei der Erstellung einer Vorlagendatei (.csv) zur Eingabe der Lab-Werte.
4.
Erstellen Sie Ihre eigenen Kalibrierungen mit dem NIR-Predictor, um Ihre NIR- und Lab-Daten für eine Kalibrierungsanforderung zu kombinieren :
siehe Video
lese Handbuch
Avvio rapido del flusso di lavoro NIR: passo dopo passo
1.
Controlla se il tuo NIR-Device Data Format è supportato direttamente (in ogni caso puoi convertirlo in JCAMP) :
NIR-Predictor supported Spectral Data File Formats
2.
Scarica il software gratuito NIR-Predictor che contiene i dati dimostrativi per poter giocare con esso per vedere se è il modo in cui vuoi analizzare i tuoi spettri NIR (non è necessaria la registrazione) :
NIR-Predictor Download
3.
Con il software di misura "NIR device" :
- misurare i campioni con il NIR, in modo da ottenere i file degli spettri,
- etichettarli con un nome di esempio corretto, in modo da sapere qual è quale,
- e determinare i valori quantitativi di riferimento con il metodo di riferimento di laboratorio.
- sono necessari almeno 60 campioni con contenuti diversi per una calibrazione minima.
- NIR-Predictor aiuta a creare un file modello (.csv) per inserire i valori di laboratorio.
4.
Creazione di calibrazioni personalizzate con NIR-Predictor per combinare i dati NIR e di laboratorio per una richiesta di calibrazione :
vedi il Video
leggi il manuale